Roche Cobas C 311 Clinical Automated Chemistry Analyzer

Contact us with any questions.

Thank you!

 Request Quote

Description

Roche Cobas C 311 Clinical Automated Chemistry Analyzer

The Cobas c 311 analyzer is a stand alone system that offers consolidated testing from a broad menu of clinical chemistry applications. This analyzer has the capacity for ion-selective electrode (ISE) determination of sodium, potassium, and chloride in serum, plasma, and urine. In addition, measurement of HbA1c levels in whole blood can also be performed, making it a truly flexible analyzer.

Includes:  

Cobas c 311 analyzer including PC with software and accessories:

Cobas Link, Bottles,

Probes,

Printer,

Touch monitor and all connections.

 

Features/Benefits

Features: 

First class performance

– Over 130 assays and applications available (including drugs of abuse testing, Therapeutic Drug Monitoring (TDM), specific proteins and whole blood HbA1c1

– Predictable turnaround time

 

Intelligent sample workflow 

– 108 sample positions with continuous random access and flexible STAT priority settings1

– Automatic determination of sample integrity  

– Automatic sample dilution and rerun capabilities

 

Unique reagent concept

– Convenient and error-free handling of cobas c packs

– No preparation or mixing required

– Economic usage with high stabilities and convenient kit sizes

 

High system reliability

– Highly reliable system based on more than 35 years of experience

– Comprehensive training and support from Roche organizations worldwide

– Automated maintenance functions 

– Low system maintenance time

Specifications

Test throughput
  • Up to 300 photometric tests/hour
  • 450 ISE tests/hour
Number of channels 42 channels and 3 ISEs
Programmable parameters 86 photometric applications, 3 ISE applications, 3 serum indices, 8 calculated results
Sample material Serum/plasma, urine, CSF, WB (HbA1c only)
Sample input/output Disk with 108 positions for samples, calibrators and controls, plus W1 position for ISE cleaning solution, W2 position for ISE activator; user-definable STAT positions
Sample barcode types NW7 (Codabar); Code 39; ITF; Code 128
System interface RS-232C serial interface
Sample database 10,000 results (routine, STAT and QC)
Calibration methods Start-up, re-calibration for photometric assays: Linear, non-linear multi points, 2 point calibration, K-factor. Up to 100 calibrators pre-programmable, storage of up to 180 curves. Preventative calibration of standby cassettes. Two K-factors can be defined for different sample types.
QC methods Individual QC and cumulative QC. Up to 100 controls pre-programmable/Preventative QC after calibration of standby cobas c packs/Auto QC (without operator intervention).
Electrical requirements 115 Volts AC; Frequency 60 Hz +/- 0.5%
Water requirements Water Type: bacteria-free, deionized water
Water consumption ~ 12 L/hr during operation
Operating conditions Ambient room temperature of 15 to 32°C; relative humidity 30 to 85% non-condensing
Physical dimensions
  • Depth: 34 in (86.4 cm)
  • Height closed: 50 in (127 cm)
  • Height open: 67 in (170.2 cm)
  • Width: 52 in (132.1 cm)
Weight 550 lbs (250 kg)

Test Menu

A1C-3 BILT3 GGT-2 MTQL TRIGL
A1MG2 C3C-2 GLUC2 MYO2 TRSF2
AAGP2 C4-2 GLUC3 NAPA2 UA2
AAT2 CA2 HAPT2 NH3L UIBC
ACET2 CARB4 HCYS NH3L2 UREAL
ALB2 CERU HDLC4 OPI2 VALP2
ALBP CHE IGA-2 OXY VANC3
ALBT2 CHE2 IGG-2 PCP β2-Microglobulin
ALP2 CHOL2 IGM-2 PHNO2
ALTL CK IRON2 PHNY2
ALTLP CO2-L ISE Chloride Electrode PHOS2
AMIK 2 COC2 ISE Potassium Electrode PPX
AMPS2 CREJ2 ISE Sodium Electrode PREA
AMY-P CREP2 KFLC PROC2
AMYL2 CRP4 LACT2 QUIN2
APOAT CRPHS LDHI2 RF-II
APOBT CRPLX LDLC3 SALI
ASLOT CYSC2 LFLC SI2
ASTL D Bili LI STFR
ASTLP D-DI LIPC THC2
B2MG DIG LPA2 THEO2
BARB ETOH2 MDN2 TOBR2
BENZ FERR4 MG2 TP2
BILT2 FRA MPA TPUC3