The Awareness Technology Stat Fax 3300 Microplate Reader is designed to read and calculate the results of in-vitro diagnostic assays. Equipped with a large graphical display, the reader provides absorbance and concentration readings at or near the available wavelength. The flexible Stat Fax 3300 is equipped with multiple, self-prompting operating modes to meet the varying demands and requirements of any facility. The modes include Absorbance Mode, Standard Mode, Rate Mode, Factor Mode, Multi-point Mode, and Index Mode. Moreover, the instrument accepts standard 12 mm round tubes as well as 1 cm square cuvettes.

Additional information

Manufacturer

Features/Benefits

  • Optical system temperature controlled to 37 C
  • Bi-chromatic with 6 filters and spaces for 2 optional additional filters (340-700nm)
  • Adapts for use with 12mm round tubes, 1cm square cuvettes, or built-in flow cell
  • Vacuum system speeds sampling and reduces carryover
  • Calibration traceable to NIST standard
  • Use the optional external keyboard for easy entry of patient names and data
  • Use Redi-Check to monitor accuracy, linearity, and repeatability the easy way
  • Peltier cell cooling available as an option

View kinetic plots in real time

Calculate kinetic results based on the entire read time OR the segment you choose.

Stat Fax 3300 stores patient data

Create a work list.
Run the tests.
Print Patient Reports.
Output data to:
– On-board 40-column
Thermal printer
– External printer
– PC and LIS.

Stat Fax 3300 storescontrol values

and of fers many QC options to assure linear kinetics and reliable results.
CreateLevey-Jennings plots for QC

Stat Fax 3300 stores calibrations and standard curves

For each run, choose a new curve or use a stored curve with options to revert to previous curve and to modify multi- point curves.

Specifications

Spectrophotometer Type Filter photometer
Optical Configuration Single beam with continuously rotating filter wheel Monochromatic
or bichromatic reading 8 filter positions
Usable Spectral Range 330 to 770 nm
System Procedures Open and by stored menu
Calculating Modes Absorbance, Single Standard, Factor Mode, Multi Standard Mode (up to 7 standards),
Multi Standard % Abs (up to 7 standards), Kinetic Mode, Index Mode
Channels 120 Open
Source of Radiation Tungsten Halogen, 10 Watt, with automatic lamp saver
Selection of Wavelength By filter
Filter Type 4-cavity interference, long-life ion beam-assisted deposition
Wavelength Accuracy +/- 3 nm
Filter Location After sample (heat absorbing filter before sample)
Filter Selection Automatic by software or via keyboard
Wavelengths 340, 405, 505, 545, 580, 630 nm supplied standard, other/additional filters optional
Half Bandwidth < 10 nm
1/100 Bandwidth 14 nm at 340 nm
False Radiant Energy Ratio < 0.001 at 340 and 405 nm
Cuvette 1 cm square, 12 cm round, flow through
Supplied Type flow through
Material 316 stainless, borosilicate windows
Geometry Cylindrical, 2.3 mm dia x 5 mm +/- 0.05 mm
Illuminated Volume 21 µl
Minimum Read Volume 250 µl
Aspiration/Purge Vacuum pump at 18 cm of Hg
Valve Silicone pinch type
Cuvette Holder Thermostatically controlled compartment at 37° C
Detector Gallium-Arsenide-Phosphide photodiode
Display Type 240 x 128 Graphic LCD w/ Backlight
Absorbance -0.5 to 3.5 (flow-through mode) -0.5 to 2.5 (tube or 1 cm cuvette)
Concentration Maximum 999,999
Kinetic Results Abs/min with resolution of 0.0002 A/min
Zero Compensation Automatic
Range -0.5 to 2.5 absorbance
Parallel Centronics/IBM-PC compatible
Serial RS-232 at 9600 baud, 8 data, 1 stop, no parity Bi-Directional
Data Input 1) 20 Key Keypad, 2) PS2 101 Keyboard (connector available at back of instrument)
Flow-through < 0.5 % at 1 absorbance, 340/630 nm
Stability Better than 0.003 A/hr monochromatic after warm-up
Better than 0.001 A/hr bichromatic after warm-up
Warm-up Time 90 seconds photometric 15 minutes for temperature compartment
Electronics Z180 Microprocessor 18 MHz 128k EEPROM 32 K bytes Nonvolatile RAM (NVRAM)
Power Auto-Switching Power Supply
Voltage source: 90 – 264 VAC
Frequency: 50/60 Hz
Power consumption: 60 watts
Installation Category: CAT II
Fuses:
2.5A/250V Fast 5-20 mm Glass Fuse,
2.0A/250V Fast 5-20 mm Ceramic Fuse,
(2) 6/10 250V Slow Blow 3AG Fuses
Dimensions and Weight 40cm (L) x 37 cm (W) x 14 cm (H) lid closed (30 cm lid opened), 6.4 kg