Beckman Coulter / Olympus AU 480 Chemistry Analyzer

Contact us with any questions.

Thank you!

Request Quote

IDE-difference

Description

Beckman Coulter / Olympus AU480 Chemistry Analyzer 

The AU480 chemistry analyzer is the ideal main analyzer for small- to medium-sized laboratories/Hospitals, with a throughput of up to 800 tests per hour with ISE and simultaneous programming for up to 63 different analytes. The Olympus AU 480 chemistry analyzer can be used as a special chemistry or STAT analyzer in any laboratory.

Robust testing capabilities

  • Broaden your test offerings with more than 120 different Beckman Coulter system reagent applications available
  • Conduct quick priority testing using the AU 480 clinical chemistry analyzer’s STAT wheel
  • Turn around critical samples fast with sample pre-dilution and auto-repeat of over-range results

Efficient reagent and consumable use

  • Minimize hands-on time and reduce daily reagent inventory tasks with ready-to-use liquid concentrated reagents
  • Experience lower total cost of ownership with low consumable requirements

Convenient operational features

  • Provide the right operator support when needed using context-sensitive help screens and instructional videos
  • Simplify operations with user-friendly graphical user interface design and touch-screen monitor
  • Reduce tech-to-tech variation and achieve consistent result quality with AU family of chemistry analyzers’ standardized processes

 

Additional information

Manufacturer

,

Features

Robust testing capabilities

  • Broaden your test offerings with more than 120 different Beckman Coulter system reagent applications available
  • Conduct quick priority testing using the analyzer’s STAT wheel
  • Turn around critical samples fast with sample pre-dilution and auto-repeat of over-range results

Efficient reagent and consumable use

  • Minimize hands-on time and reduce daily reagent inventory tasks with ready-to-use liquid concentrated reagents
  • Experience lower total cost of ownership with low consumable requirements

Convenient operational features

  • Provide the right operator support when needed using context-sensitive help screens and instructional videos
  • Simplify operations with user-friendly graphical user interface design and touch-screen monitor
  • Reduce tech-to-tech variation and achieve consistent result quality with AU family of chemistry analyzers’ standardized processes

Specifications

Analytical System:          Fully automated, random access chemistry system with STAT capability

Analytical Principle:        Spectrophotometry and potentiometry

Analytical Types:           Endpoint, rate, fixed point and indirect ISE

Analytical Methods:        Colorimetry, turbidimetry, latex agglutination, homogeneous EIA, indirect ISE

Simultaneously Processed Analytes:  60 photometric tests + 3 ISE, 120 pre-programmed onboard tests

Throughput:                  400 photometric tests/hour; maximum of 800 with ISE

ISE Throughput:             200 samples/hr. Maximum tests/hr: 600 if ISE only

Sample Types:                Serum, plasma, urine, other

Sample Feeder:               Racks with 10 samples each (barcodes on primary tubes and on racks); capacity of 80 samples; continuous loading

Sample Tubes:                Primary and secondary tubes; diameter from 11.5 to 16 mm; height from 55 to 102 mm, nestled micro sample cups

STAT samples:                Up to 22 positions for STAT samples via cooled turntable. Easy-to-operate interrupts between tests. Auto-repeat run capability.

Sample Volume:              1–25 μl in 0.1 μl steps (1-25 μl for repeats)

Reagent Capacity:           76 positions for (R1+R2, detergent position). Holds 15, 30 and 60ml bottles.

Reagent Storage:            Refrigerated (4–12°C)

Reagent Volume:             R1: 10–250 μl; R2: 10–250 μl (in 1 μl increment)

Total Reaction Volume:    90–350 μl

Reaction Cuvette:           Permanent glass cuvettes

Reaction Time:               Up to 8 minutes, 37.5 seconds

Reaction Incubation:       37°C, dry bath

Photometry System:       Direct assay through the reaction cuvette (0–3.0 OD) mono and bichromatic measurements possible

Wavelength:                  13 different wavelengths between 340–800 nm

Calibration:                    Auto calibration, cooled calibrator positions; master calibration established by two-dimensional barcode

Quality Control:              Auto QC, cooled QC positions

Rerun:                           1. Auto rerun and manual rerun available

2. Automatic sample increase, decrease or normal repeat

3. Reflex testing

Sample Pre-dilution:       Dilution ratio: Fixed to 3, 5, 10, 15, 20, 25, 50, 76 or 100

Test Requisition:            Individual and profile test requisition via online, mouse, function keys and touch-screen

Sample Integrity:           Lipaemia, Haemolysis and Icterus analysis. Sample clot detection and probe crash protection.

Online:                         RS232C. Full uni- and bidirectional communication possible.

Software:                      Windows 7®

Data Storage:               Sample number 100,000 samples. Reaction Data 200,000 tests.

Dimensions (W x H x D):     ANL: 1,450 x 1,205 x 760

Power Supply:                    100–240v; 60 Hz/ < 3.5 kVA

Test Menu

Acetaminophen D-Dimer Phosphorus
ALT Digoxin Prealbumin
Albumin Disopyramide Primidone
Alcohol Ecstasy Procainamide
ALP Ethosuximide Propoxyphene
a1-acid glycoprotein Ferritin Quinidine
a1-antitrypsin Fructosamine Rheumatoid Factor
Amikacin GGT Salicylate
Ammonia Gentamicin T-Uptake
Amphetamine Glucose Theophylline
Amylase Haptoglobin Thyroxin, Total
Anti-strepolysin O HDL Cholesterol, Direct Tobramycin
Apolipoprotein A1 Hemoglobin A1c Transferrin
Apolipoprotein B IgA Triglyceride
AST IgG UIBC
Barbiturate IgM Urea Nitrogen
Benzodiazepine Iron Uric Acid
B2 Microglobulin Lactate Dehydrogenase Valproic Acid
Bicarbonate Lactate Vancomycin
Bilirubin, Direct LDL Cholesterol, Direct
Bilirubin, Total Lidocaine
CRP, hs Lipase
CRP Lithium
C3 Complement LSD
C4 Complement Magnesium
Caffeine Methadone
Calcium Methaqualone
Carbamazepine Methotrexate
Ceruloplasmin Microalbumin
Chloride N-Acetylprocainamide
Cholinesterase Opiate Metabolite
Cocaine Metabolite Oxycodone
CK-MB Phencyclidine (PCP)
CK-NAC Phenobarbital
Creatinine Phenytoin